OMD- 24 SeriesAlarm 1 operating delay for annunciation
Alarm 2 operating delay for control
1 - 10 sec
Alarm contacts
2 independently adjustable switchover contacts
Alarm contact operation mode
De-energized in Alarm State
Alarm contact ratings
3A, 240V each
Alarm output
0 - 20 mA or 4 - 20 mA, for 0-30ppm reversible, ext. Load<150Ω
Display
Graphic Display
Ambient temperature
+1°C to +55°C
Class of protection
IP 65
Dimension Computer Unit
185 x 210 x 65 mm (WxHxD)
Dimension Measuring Cell
140 x 160 x 65 mm (WxHxD)
Weight
2,5 kgsProperty
Value
Range (measurement & indication)
0 - 30 ppm, Trend up to 50ppm
Range (IMO specification)
0 - 15 ppm (according IMO regulations)
Accuracy
better than IMO MEPC.107(49)
Electrical power supply
24V, AC or DC, +/-10%
Electrical power consumption
< 5 VA
Sample
Oily water from separator outlet, 0,1 - 4 l/min
Sample temperature range
+1°C to +65°C
Sample connections
R 1/4" Female
Alarm 1 & 2 Set Points
1 - 15ppm (adjustable) works adjustment 15ppm